Goldway G30E

Features

  • Advanced functionalities
    To accommodate different clinical environments, the G30E delivers real-time data about cardiac and respiratory conditions. In addition to basic monitoring parameters, such as Philips ST/AR 3- and 5-lead ECG/ RESP, NBP, 2 temperatures, and Philips SpO₂/PR, ST and arrhythmia monitoring, it also provides 2-channel invasive blood pressure and microstream etCO₂ monitoring as options.
  • Ease of use
    The monitor has a 10.4” display with flexible display modes, such as big number display mode, minitrend, OxyCRG, as well as auto waveform channel height adjustment function to fit various clinical settings. Intuitive menus, fixed key access, bright colour screen and optional touchscreen were all designed for maximum ease of use.
  • Networking and data management
    The G30E comes standard with LAN networking capabilities for facilities looking to create a network with a bi-directional communication with Philips Efficia CMS200 central monitoring system or require data output in HL7 format.
  • Compact and cost-effective
    Built according to Philips exacting quality standards, the G30E patient monitor offers market-proven essential parameters in a compact, cost-effective monitor to provide accurate measurement results within your tight budget. The alarm latch function helps reduce alarm fatigue to form a silent care environment.
  • Works with Efficia consumables
    The G30E patient monitor is designed to be compatible with Philips Efficia consumables, which are also used on Philips Efficia CM series bedside patient monitors, to lower your total cost of ownership by sharing Philips consumables across different care settings.

Specifications

  • ECG
    Lead type
    3-lead or 5-lead selectable
  • Lead selection, 3-lead
    I, II, III
    Lead selection, 5-lead
    I, II, III, aVR, aVL, aVF, V
  • HR measurement
    Adult / pediatric: 15 bpm – 300 bpm Neonatal: 15 bpm – 300 bpm Step: 1 bpm
  • HR accuracy
    ±5 bpm or ±1 %, whichever is greater
  • Arrhythmia
    21 calls
  • ST measurement range
    -2.5 mV – +2.5 mV
  • Waveform speed
    6.25 mm/s,12.5 mm/s, 25 mm/s, 50 mm/s
  • Gain selection
    x 1/4, x 1/2, x 1, x 2, x 4, Auto
  • Pacemaker detection
    Yes
  • Protection
    Against electrosurgical interference and defibrillation
  • Non-invasive blood pressure (NBP)
    Method
    Automatic oscillometric
  • Parameters
    SYS/DIA/MAP/Pulse Rate
  • Operating mode
    Manual/Auto/STAT
  • Auto measure intervals
    3, 5, 10, 15, 30, 60, 90, 120, 240, 480 minutes
  • Units
    mmHg/kPa
  • Cuff pressure range, adult / pediatric
    0 – 300 mmHg ( 0 kPa – 40.0 kPa)
  • Cuff pressure range, neonate
    0 – 150 mmHg ( 0 kPa – 20.0 kPa)
  • Systolic measurement range, adult/ pediatric
    30 mmHg – 254 mmHg (4.0 kPa – 33.9 kPa)
    Systolic measurement range, neonate
    30 mmHg – 135 mmHg (4.0 kPa – 18.0 kPa)
  • Diastolic measurement range, adult / pediatric
    10 mmHg – 220 mmHg (1.3 kPa – 29.3 kPa)
    Diastolic measurement range, neonate
    10 mmHg – 110 mmHg (1.3 kPa – 14.7 kPa)
  • MAP measurement range, adult / pediatric
    20 mmHg – 235 mmHg (2.7 kPa – 31.3 kPa)
    MAP measurement range, neonate
    20 mmHg – 125mmHg (2.7 kPa – 16.7kPa)
  • Accuracy, maximum standard deviation
    8 mmHg (1.1 kPa)
  • Accuracy, mean error
    ±5 mmHg (±0.7 kPa)
  • Pulse rate range
    40 bpm – 240 bpm
  • Pulse rate accuracy
    ±5bpm or ±5%, whichever is greater
  • SpO2 (Philips SpO2 technology)
    Measurement range
    0 – 100 %
    Accuracy
    70 % – 100 %, depends on sensor; 0 – 69 %, unspecified
  • Waveform speed
    6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s
  • Pulse rate range
    30 bpm – 300 bpm
  • Pulse rate accuracy
    ±2 bpm or ±2%, whichever is greater
  • Perfusion index
    Yes
  • Respiration (RESP)
    Measurement method
    Thoracic impedance
    Measurement range
    0 rpm – 150 rpm
    Accuracy
    0rpm – 120rpm: ±1 rpm; 121rpm – 150rpm: ±2 rpm
  • Waveform speed
    6.25 mm/s,12.5 mm/s, 25 mm/s, 50 mm/s Accuracy undefined
  • Gain selection
    x 1/2, x 1, x 2
  • Apnea alarm
    Yes, 10 – 60 s
  • Work mode
    Auto or manual (threshold)
  • RESP lead
    RA-LL
  • Temperature (TEMP, 2-channel)
  • Measurement range
    0 – 50℃ (32℉ – 122℉)
  • Resolution
    0.1℃ (0.2℉)
  • Accuracy
    ±0.1℃ (±0.2℉) (without sensor)
  • Units
    ℃ / ℉
  • Minimum measuring time
    Attachable skin surface probe: less than 10 minutes Esophageal/rectal probe: less than 2 minutes
  • Capnopgraphy (etCO2, optional)
  • Method
  • Microstream, infrared absorption (Oridion)
  • Parameters
    awRR, etCO₂, inCO₂
  • Measurement range
    0 – 99 mmHg (0 kPa – 13.2 kPa)
  • Accuracy+C781, 0 – 38 mmHg (0 kPa – 5.1 kPa)
    ±2 mmHg (±0.28 kPa)
  • Accuracy, 39 mmHg – 99 mmHg (5.2 kPa – 13.2 kPa)
    ± (5% of reading + 0.08% for every 1 mmHg (0.13 kPa) above 38 mmHg (5.1kPa)
    awRR range
    0-150 rpm
    awRR accuracy, 3 rpm – 70 rpm
    ±1 rpm
    awRR accuracy, 71 rpm – 120 rpm
    ±2 rpm
    awRR accuracy, 121 rpm – 150 rpm
    ±3 rpm
  • Invasive Blood Pressure
    Accuracy
    ±4 mmHg (0.53 kPa) or ±4 % of reading, whichever is greater
    Invasive blood pressure (IBP, 2-channel, optional)
  • Parameters
    SYS/DIA/MAP/Pulse Rate
  • Measurement range
    -50 – 400 mmHg (-6.6 kPa – 53.3 kPa)
  • Waveform speed
    6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s
  • Resolution
    1 mmHg (0.13 kPa)
  • Units
    mmHg/kPa/cmH₂O
  • Transducer sensitivity
    5 uV/V/mmHg
  • Pulse rate range
    30 bpm – 254 bpm
  • Pulse rate accuracy
    ±2bpm or ±2%, whichever is greater
  • Sampling rate
    62.5 samples per second
  • Recorder (optional)
    Type
    Built-in thermal recorder, 3 channels
    Print mode
    Real-time recording, timed recording, alarm-activated recording
    Print speed
    Auto, 12.5 mm/s, 25 mm/s, 50 mm/s
  • Display
    Type
    Color TFT LCD, 4:3
  • Size
    10.4+611:631″
  • Resolution
    800 x 600 pixels
  • Channels
    Max 11 channels (auto waveform channel height adjustment)
  • Power supply
    AC power input
    100 – 240 VAC, 50/60 Hz
  • Battery
    Rechargeable Li-Ion battery Maximum 2.5 hour continuous monitoring (1 battery)
  • Weight and dimensions
    Net weight
    Approx. 3.5 kg without battery
  • Dimensions (w x d x h)
    316 mm x 265 mm x 131 mm (12.5 in. x 10.5 in. x 5.2 in.)
  • Safety approval and quality system
    Safety standard
    IEC 60601-1-2:2014
    Electromagnetic environment
    IEC 60601-1-2:2000/A1:2014
  • Quality system
    ISO9001 & ISO13485 certified CE marking according to the Council Directive 93/42/EEC
    Water proof
    IPX1
  • Data review and storage
    Data review and storage
    1600- NBP data, 1200 hr trend graph, 120 min full disclosure ECG 1, 200 – alarm & arrhythmia events
  • External interface (standard)
    External interface
    ECG analog output VGA output Nurse call LAN USB
Enquiry Now

Description

Goldway G30E
Bedside patient monitor

Clinicians need to respond swiftly to rapid changes in their patient’s condition. Having a highly reliable, versatile patient monitoring solution on hand saves precious time in determining the best course of care. Philips Goldway G30E is your choice for a cost-effective, easy to learn, pre-configured bedside monitor that provides reliable measurements (Philips ST/AR ECG and FAST-SpO₂).