Features
- Advanced functionalities
To accommodate different clinical environments, the G30E delivers real-time data about cardiac and respiratory conditions. In addition to basic monitoring parameters, such as Philips ST/AR 3- and 5-lead ECG/ RESP, NBP, 2 temperatures, and Philips SpO₂/PR, ST and arrhythmia monitoring, it also provides 2-channel invasive blood pressure and microstream etCO₂ monitoring as options.- Ease of use
The monitor has a 10.4” display with flexible display modes, such as big number display mode, minitrend, OxyCRG, as well as auto waveform channel height adjustment function to fit various clinical settings. Intuitive menus, fixed key access, bright colour screen and optional touchscreen were all designed for maximum ease of use.- Networking and data management
The G30E comes standard with LAN networking capabilities for facilities looking to create a network with a bi-directional communication with Philips Efficia CMS200 central monitoring system or require data output in HL7 format.- Compact and cost-effective
Built according to Philips exacting quality standards, the G30E patient monitor offers market-proven essential parameters in a compact, cost-effective monitor to provide accurate measurement results within your tight budget. The alarm latch function helps reduce alarm fatigue to form a silent care environment.- Works with Efficia consumables
The G30E patient monitor is designed to be compatible with Philips Efficia consumables, which are also used on Philips Efficia CM series bedside patient monitors, to lower your total cost of ownership by sharing Philips consumables across different care settings.Specifications
- ECG
Lead type
3-lead or 5-lead selectable- Lead selection, 3-lead
I, II, III
Lead selection, 5-lead
I, II, III, aVR, aVL, aVF, V- HR measurement
Adult / pediatric: 15 bpm – 300 bpm Neonatal: 15 bpm – 300 bpm Step: 1 bpm- HR accuracy
±5 bpm or ±1 %, whichever is greater- Arrhythmia
21 calls- ST measurement range
-2.5 mV – +2.5 mV- Waveform speed
6.25 mm/s,12.5 mm/s, 25 mm/s, 50 mm/s- Gain selection
x 1/4, x 1/2, x 1, x 2, x 4, Auto- Pacemaker detection
Yes- Protection
Against electrosurgical interference and defibrillation- Non-invasive blood pressure (NBP)
Method
Automatic oscillometric- Parameters
SYS/DIA/MAP/Pulse Rate- Operating mode
Manual/Auto/STAT- Auto measure intervals
3, 5, 10, 15, 30, 60, 90, 120, 240, 480 minutes- Units
mmHg/kPa- Cuff pressure range, adult / pediatric
0 – 300 mmHg ( 0 kPa – 40.0 kPa)- Cuff pressure range, neonate
0 – 150 mmHg ( 0 kPa – 20.0 kPa)- Systolic measurement range, adult/ pediatric
30 mmHg – 254 mmHg (4.0 kPa – 33.9 kPa)
Systolic measurement range, neonate
30 mmHg – 135 mmHg (4.0 kPa – 18.0 kPa)- Diastolic measurement range, adult / pediatric
10 mmHg – 220 mmHg (1.3 kPa – 29.3 kPa)
Diastolic measurement range, neonate
10 mmHg – 110 mmHg (1.3 kPa – 14.7 kPa)- MAP measurement range, adult / pediatric
20 mmHg – 235 mmHg (2.7 kPa – 31.3 kPa)
MAP measurement range, neonate
20 mmHg – 125mmHg (2.7 kPa – 16.7kPa)- Accuracy, maximum standard deviation
8 mmHg (1.1 kPa)- Accuracy, mean error
±5 mmHg (±0.7 kPa)- Pulse rate range
40 bpm – 240 bpm- Pulse rate accuracy
±5bpm or ±5%, whichever is greater- SpO2 (Philips SpO2 technology)
Measurement range
0 – 100 %
Accuracy
70 % – 100 %, depends on sensor; 0 – 69 %, unspecified- Waveform speed
6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s- Pulse rate range
30 bpm – 300 bpm- Pulse rate accuracy
±2 bpm or ±2%, whichever is greater- Perfusion index
Yes- Respiration (RESP)
Measurement method
Thoracic impedance
Measurement range
0 rpm – 150 rpm
Accuracy
0rpm – 120rpm: ±1 rpm; 121rpm – 150rpm: ±2 rpm- Waveform speed
6.25 mm/s,12.5 mm/s, 25 mm/s, 50 mm/s Accuracy undefined- Gain selection
x 1/2, x 1, x 2- Apnea alarm
Yes, 10 – 60 s- Work mode
Auto or manual (threshold)- RESP lead
RA-LL- Temperature (TEMP, 2-channel)
- Measurement range
0 – 50℃ (32℉ – 122℉)- Resolution
0.1℃ (0.2℉)- Accuracy
±0.1℃ (±0.2℉) (without sensor)- Units
℃ / ℉- Minimum measuring time
Attachable skin surface probe: less than 10 minutes Esophageal/rectal probe: less than 2 minutes- Capnopgraphy (etCO2, optional)
- Method
- Microstream, infrared absorption (Oridion)
- Parameters
awRR, etCO₂, inCO₂- Measurement range
0 – 99 mmHg (0 kPa – 13.2 kPa)- Accuracy+C781, 0 – 38 mmHg (0 kPa – 5.1 kPa)
±2 mmHg (±0.28 kPa)- Accuracy, 39 mmHg – 99 mmHg (5.2 kPa – 13.2 kPa)
± (5% of reading + 0.08% for every 1 mmHg (0.13 kPa) above 38 mmHg (5.1kPa)
awRR range
0-150 rpm
awRR accuracy, 3 rpm – 70 rpm
±1 rpm
awRR accuracy, 71 rpm – 120 rpm
±2 rpm
awRR accuracy, 121 rpm – 150 rpm
±3 rpm- Invasive Blood Pressure
Accuracy
±4 mmHg (0.53 kPa) or ±4 % of reading, whichever is greater
Invasive blood pressure (IBP, 2-channel, optional)- Parameters
SYS/DIA/MAP/Pulse Rate- Measurement range
-50 – 400 mmHg (-6.6 kPa – 53.3 kPa)- Waveform speed
6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s- Resolution
1 mmHg (0.13 kPa)- Units
mmHg/kPa/cmH₂O- Transducer sensitivity
5 uV/V/mmHg- Pulse rate range
30 bpm – 254 bpm- Pulse rate accuracy
±2bpm or ±2%, whichever is greater- Sampling rate
62.5 samples per second- Recorder (optional)
Type
Built-in thermal recorder, 3 channels
Print mode
Real-time recording, timed recording, alarm-activated recording
Print speed
Auto, 12.5 mm/s, 25 mm/s, 50 mm/s- Display
Type
Color TFT LCD, 4:3- Size
10.4+611:631″- Resolution
800 x 600 pixels- Channels
Max 11 channels (auto waveform channel height adjustment)- Power supply
AC power input
100 – 240 VAC, 50/60 Hz- Battery
Rechargeable Li-Ion battery Maximum 2.5 hour continuous monitoring (1 battery)- Weight and dimensions
Net weight
Approx. 3.5 kg without battery- Dimensions (w x d x h)
316 mm x 265 mm x 131 mm (12.5 in. x 10.5 in. x 5.2 in.)- Safety approval and quality system
Safety standard
IEC 60601-1-2:2014
Electromagnetic environment
IEC 60601-1-2:2000/A1:2014- Quality system
ISO9001 & ISO13485 certified CE marking according to the Council Directive 93/42/EEC
Water proof
IPX1- Data review and storage
Data review and storage
1600- NBP data, 1200 hr trend graph, 120 min full disclosure ECG 1, 200 – alarm & arrhythmia events- External interface (standard)
External interface
ECG analog output VGA output Nurse call LAN USB
Goldway G30E
Description
Goldway G30E
Bedside patient monitorClinicians need to respond swiftly to rapid changes in their patient’s condition. Having a highly reliable, versatile patient monitoring solution on hand saves precious time in determining the best course of care. Philips Goldway G30E is your choice for a cost-effective, easy to learn, pre-configured bedside monitor that provides reliable measurements (Philips ST/AR ECG and FAST-SpO₂).